Mobile Rapid Transfuser
One knob, one button, both hands free. A handheld, battery-powered pump built to deliver blood and fluid before the hospital.
The pump head sits outside. Everything else stays in.
Pump head
A three-roller peristaltic head mounts on the outside of the rear wall. The pump segment loads, and comes out, without opening the housing.
Controls
One rate knob marked 1, 2 and 3 (×100 mL/min) and one latching power button. Its ring lights green when the pump is powered.
Power
A 12 V, 67.2 Wh lithium-ion pack with an integral protection circuit. It charges in place at 12.6 V / 1 A through the front-panel jack.
Fluid path
A standard sterile blood set with a PharMed® BPT pump segment, 4.8 mm ID × 8.0 mm OD. Only the single-use set touches the fluid.
One 500 mL unit, three speeds.
Minutes and seconds to deliver 500 mL at the nominal panel settings. Loaded-flow testing is in progress.1
Four steps. No tools.
Spike and prime a standard blood set with the MRT pump segment in line.
Open the pump-head cover, lay the segment in, close it. No tools; the housing stays closed.
Connect the patient line. Turn the knob to 1, 2 or 3.
Press the button. The ring lights green, the fluid moves, both hands are free.
Verified specifications
Every figure comes from a manufacturer's label or datasheet, or was measured on the prototype.
- Pump
- Three-roller peristaltic head, rated 400–500 mL/min1
- Pump segment
- PharMed® BPT, 4.8 mm ID × 8.0 mm OD, in a sterile single-use set
- Motor
- 12 V brushed DC
- Controls
- 100 / 200 / 300 mL/min settings on a continuous knob; 16 mm stainless latching pushbutton with green LED ring
- Battery
- 12 V lithium-ion, 5,600 mAh / 67.2 Wh, 3S2P, integral protection circuit
- Charging
- 12.6 V / 1 A through a front-panel barrel jack
- Size
- 80 mm wide × 110 mm tall × 150 mm long (3.1 × 4.3 × 5.9 in)
- Weight
- 696 g (1.53 lb) with battery
- Housing
- 3D-printed prototype; pump head external, electronics inside
- Patent
- U.S. provisional application 64/191,753, filed November 4, 2026
Where the MRT stands.
Prototype. Not FDA cleared. Not for clinical use. Specifications describe the prototype as built and may change.