AMS, Auto Medic Systems
The MRT

Mobile Rapid Transfuser

One knob, one button, both hands free. A handheld, battery-powered pump built to deliver blood and fluid before the hospital.

U.S. patent pending · Prototype, not FDA cleared
The MRT resting in an open hand, tubing connected, power ring lit green
696 g
with the battery installed · 1.53 lb
100 · 200 · 300
mL/min panel settings, nominal1
67.2 Wh
12 V lithium-ion pack · 5,600 mAh
80 × 110 × 150
mm wide × tall × long · 3.1 × 4.3 × 5.9 in
Front panel of the MRT: rate knob marked 1, 2 and 3, charge port and power button
Design

The pump head sits outside. Everything else stays in.

Pump head

A three-roller peristaltic head mounts on the outside of the rear wall. The pump segment loads, and comes out, without opening the housing.

Controls

One rate knob marked 1, 2 and 3 (×100 mL/min) and one latching power button. Its ring lights green when the pump is powered.

Power

A 12 V, 67.2 Wh lithium-ion pack with an integral protection circuit. It charges in place at 12.6 V / 1 A through the front-panel jack.

Fluid path

A standard sterile blood set with a PharMed® BPT pump segment, 4.8 mm ID × 8.0 mm OD. Only the single-use set touches the fluid.

At the panel settings

One 500 mL unit, three speeds.

Setting 1
100 mL/min
5:00
Setting 2
200 mL/min
2:30
Setting 3
300 mL/min
1:40

Minutes and seconds to deliver 500 mL at the nominal panel settings. Loaded-flow testing is in progress.1

The MRT on an ambulance counter beside a unit of blood
How it works

Four steps. No tools.

1

Spike and prime a standard blood set with the MRT pump segment in line.

2

Open the pump-head cover, lay the segment in, close it. No tools; the housing stays closed.

3

Connect the patient line. Turn the knob to 1, 2 or 3.

4

Press the button. The ring lights green, the fluid moves, both hands are free.

Specifications

Verified specifications

Every figure comes from a manufacturer's label or datasheet, or was measured on the prototype.

Dimension drawing of the MRT with front, three-quarter, rear and side views: 80 mm wide, 110 mm tall, 150 mm long
80 mm wide × 110 mm tall × 150 mm long.
Pump
Three-roller peristaltic head, rated 400–500 mL/min1
Pump segment
PharMed® BPT, 4.8 mm ID × 8.0 mm OD, in a sterile single-use set
Motor
12 V brushed DC
Controls
100 / 200 / 300 mL/min settings on a continuous knob; 16 mm stainless latching pushbutton with green LED ring
Battery
12 V lithium-ion, 5,600 mAh / 67.2 Wh, 3S2P, integral protection circuit
Charging
12.6 V / 1 A through a front-panel barrel jack
Size
80 mm wide × 110 mm tall × 150 mm long (3.1 × 4.3 × 5.9 in)
Weight
696 g (1.53 lb) with battery
Housing
3D-printed prototype; pump head external, electronics inside
Patent
U.S. provisional application 64/191,753, filed November 4, 2026
Status

Where the MRT stands.

Done
Prototype built and operated
Done
U.S. provisional patent filed November 4, 2026
Next
Loaded-flow and hemolysis testing
Target
FDA Class II, 510(k)

Prototype. Not FDA cleared. Not for clinical use. Specifications describe the prototype as built and may change.

Ready to talk?

Ask for the product brief, or tell us where you carry blood.

info@automedicsystems.com 800.932.3588